University of Pennsylvania
Radiation Safety Office
Veterinary Energized Equipment
User's Guide
Outline
I. Proper Operating Procedures
A. Radiographic Units
B. Fluoroscopic Units
C. Mobile Units
D. Dental Units
E. Therapeutic X-Ray Units
II. Techniques of External Radiation Protection
III. Shielding
A. Personnel Shielding
B. Structural Shielding
IV. Pregnant Worker Policy
V. Badging and Dosimetry Policy
A. Personnel Dosimeter Policy
B. Dosimeter Placement
C. Dose Reports
D. Pick-up and Drop-off of Dosimeters
VI. Radiation Survey of Energized Equipment
VII. Diagnostic Equipment
A. Registration of Energized Units
B. Acquisition of New Energized Units
C. Disposal of Energized Units
VIII. Important Phone Number
I. Proper Operating Procedures
A. Proper Operating Procedures for Radiographic Units
1. Limit the X-ray primary beam limited to the smallest area
possible consistent with the objectives of the clinical examination.
2. Align the X-ray beam properly with the animal and the image receptor.
3. Remain behind a protective barrier (i.e., a leaded glass wall, a leaded
door, etc.) during the entire radiographic exposure.
4. Whenever possible, use restraining, supporting, or positioning devices
for the animal, or use film for radiation exposures [PA 223.13.d].
5. State of Pennsylvania Rules and Regulations # 223.13.e:
"No Individual may be regularly employed to hold or support
animals or hold film or hold the x-ray tube head during radiation exposures.
Occupationally exposed individuals may not perform this service except in
cases in which no other method is available. An individual holding or supporting
an animal or film during radiation exposure shall wear protective gloves
and apron having a lead equivalent of not less than 0.5 millimeter and shall
be positioned so that no part of that individual's body will be struck by
the useful beam. The exposure of an occupationally exposed individual used
for this purpose shall be monitored."
6. Provide protective garments (lead aprons/shielding) for everyone whose
presence is necessary during the radiographic exposure.
B. Proper Operating Procedures for Fluoroscopic Units
1. Only persons required for a fluoroscopic procedure should be in the room
during the procedure.
2. As in a radiographic procedure, use the smallest possible beam area,
thereby reducing the scatter radiation to personnel.
3. Fluoroscopic doses can also be minimized by reduction in the fluoroscopic
time used. Perform quarterly outputs for Entrance Skin Exposure (ESE) rates
for all fluoroscopes. Post the ESE Rates in the fluoroscopy room for reference
by the physician/radiologist.
4. Use the timing device to indicate a preset time to serve as a reminder
to keep it as short as possible. According to state regulations, the predetermined
time should not exceed 5 minutes (PA 221.38).
5. Use the shortest possible distance from the image intensifier to the
animal. For fluoroscopes equipped with AEC, AEC operations provide for automatic
compensation, so that when longer distances are used, a higher radiation
dose is given .
6. Thyroid shield, leaded gloves, and glasses should be available to the
fluoroscopist to reduce exposure to thyroid, the extremities and eyes.
C. Proper Operating Procedures for Mobile Energized Units
If proper care is not taken, mobile equipment has a greater potential than
standard equipment for unnecessary radiation exposure of personnel. In mobile
units, variations can occur in exposures due to line voltage fluctuations,
which can affect the output of the machine. Rechargeable units are capacitor
discharge units are not affected.
a. Stand well away from the useful beam and the animal during
the radiographic exposures. Distance is often the best possible protection
from radiation.
b. Individuals, other than those whose presence is necessary to conduct
the x-ray procedures, must be outside the x-ray room or, for field procedures,
shall stand at least 5 meters away from the x-ray tube and from the animal.
c. Operators should wear protective garments with lead equivalent of at
least 0.5 millimeter.
D. Proper Operating Procedures for Dental Units
1. Only persons required for radiographic procedure should be
in the radiographic room during exposure. Everyone in the room must be adequately
protected by protective garments.
2. Align the x-ray beam and film very carefully with the area to be radiographed.
3. Stand behind a protective barrier and observe the animal during the dental
exposure.
4. The operator or the assistant must not hold the film in place during
the exposure. Film holder devices must be used to hold the film in the place.
E. Proper Operating Procedures for Therapeutic X-Ray Units
1. Secure therapeutic x-ray units to prevent unauthorized use
whenever the system is unattended.
2. Use mechanical supporting or restraining devices to hold the animal in
position for radiation therapy .
3. The tube housing assembly must not be held during exposures.
4. No one may be in treatment room during irradiation unless protected by
a shielded both.
5. Check interlocks, on-off beam control mechanisms, safety, and warning
devices and at least once in a calendar year and service them appropriately.
II. Techniques of External Radiation Protection
Radiation exposure levels can be controlled via four basic methods:
1. Maximization of distance from radiation source.
2. Minimization of radiation exposure time.
3. Proper shielding of radiation source.
4. Proper shielding for personnel.
III. Shielding
A. Personnel Shielding:
For personnel who remain in the room during examinations, it
is important that they be protected by proper shielding, such as the following:
a. A leaded apron must be worn any time personnel are in a room
during an exposure.
b. Thyroid shields: anyone likely to be exposed to high levels of scattered
radiation to the thyroid during any procedure should wear this shield.
c. Leaded glasses: in fluoroscopy, the lens of the eye can often be exposed
to high levels of scattered radiation.
The use of leaded glasses can greatly reduce this exposure.
d. Leaded gloves: anyone who must have his/her hand near the primary beam
(as in the case where no
other means is available to immobilize an animal ) should wear leaded gloves
to reduce exposure to the extremities.
B. Structural Shielding:
Each radiographic room has been designed with sufficient shielding
in the walls to provide
protection to anyone on the outside of the room. It is important that the
integrity of the
shielded walls not be tampered with. Therefore, if any personnel notice
structural changes,
such as holes drilled into walls, Radiation Safety should be notified as
soon as possible.
IV. Pregnant Worker Policy
An employee should contact the Radiation Safety Office either
directly or through the supervisor when she knows or suspects that she is
pregnant. If for personal reasons an employee does not wish to disclose
pregnancy to her supervisor, confidential disclosures can be made directly
to the Radiation Safety Office.
V. Badging and Dosimetry Policy
A. Personnel Dosimeter Policy
Personnel monitoring is the method used by the Radiation Safety
Office (RSO) to identify inadequate or improper radiation safety practices
and potentially serious radiation exposure situations. The RSO will issue
proper personnel dosimeters when evaluation of equipment reveals the radiation
dose to personnel could potentially be larger than ALARA limits per calendar
quarter to the whole body, (125 mrem).
Radiation workers are monitored by the University of Pennsylvania Radiation
Safety Office. All radiation workers must conscientiously wear the radiation
monitoring devices provided by the RSO. These devices may include the following:
a. A Film Badge to monitor doses to the whole body
b. A Ring Dosimeter to monitor extremity doses
c. A TLD badge for whole body measurements.
B. Dosimeter Placement
1. Interpretation of the measured dose is dependent upon the
placement of the dosimeter. It is therefore important that all personnel
wear their dosimeters correctly. The following list indicates where the
dosimeters are to be worn:
a. Film Badge with no TLD Badge: wear the film badge above any
protective clothing at collar level.
b. Film Badge with a TLD Badge: wear the film badge under the lead apron
and the TLD above the apron at collar level.
c. Ring Dosimeter: wear ring dosimeters so that the employee's name is facing
outwards.
2. Do not expose personnel monitoring devices to extreme heat or humidity.
They are screened monthly. If any dosimeter has received a dose higher than
the values shown below, the employee will be notified and the reason for
the high reading will be investigated. Measures will be taken to keep radiation
doses below these limits whenever possible.
Whole Body: 125 mrem/calendar quarter
Extremities: 1875 mrem/calendar quarter
It is very important that all personnel promptly return all monitoring devices
every month.
C. Dose Reports
The Radiation Safety Office sends dose summary reports on a
monthly and an annual basis . These reports should be posted in a conspicuous
location in the department.
D. Pick-up and drop off of dosimeters
The Radiation Safety Office delivers new dosimeters to the veterinary
school on the last Wednesday of each month. The Veterinary School must deliver
the badges for New Bolton Center on the following day. The old dosimeters
must be returned to the veterinary school designated personnel by the following
Monday. The Radiation Safety Office will pick these by the following Wednesday
for analysis.
VI. Radiation Survey of Equipment
Radiation survey of equipment checks the radiation safety characteristics
of the machine comply with FDA and PA state regulations.
All safety problems observed by personnel, including conspicuous problems
with energized equipment or with shielded rooms should be reported immediately
to the Radiation Safety Office.
VII. Diagnostic Equipment
A. Registration of Diagnostic Units
According to Section 216.2 of 25 PA Code, Chapter 216, all radiation
producing equipment must be registered with the Department of Environmental
Resources. This registration is performed by the RSO on an annual basis.
The RSO maintains a listing of all units currently registered.
B. Acquisition of New Diagnostic Units
It is the responsibility of clinical personnel to notify the
RSO upon acquisition of any new diagnostic equipment. A radiation safety
survey is conducted by authorized RSO personnel on all new units prior to
their use.
C. Disposal of Diagnostic Units
It is the responsibility of clinical personnel to notify the RSO of any
diagnostic equipment intended for disposal. The RSO will work in cooperation
with the Environmental Health & Safety Office to ensure safe disposal
of diagnostic units.
VIII. Important Phone Number
Radiation Safety Office, University of Pennsylvania: 898-7187