University of Pennsylvania
Radiation Safety Office
CHOP Radiation Safety Manual
Diagnostic Energized Equipment
Outline
I. Proper Operating Procedures
A. Radiographic Units
B. Fluoroscopic Units
C. Mobile Diagnostic Units
D. Computed Tomography Units
E.
Dental units
II. Techniques of External Radiation Protection
III. Shielding
A. Patient Shielding
B. Personnel Shielding
C. Structural Shielding
IV. Pregnant Patient & Pregnant Worker Policy
V. Badging & Dosimetry Policy
A. Personnel Dosimeter Policy
B. Dosimeter Placement
C. Dose Reports
D. Pick-up & Drop-off of Dosimeters
VI. Quality Assurance Program
A. Reject-Repeat Analysis Program
B. Film Processors
C. Radiographic X-Ray Units
D. Fluoroscopic X-Ray Units
E. Computed Tomographic X-Ray Units
F. Dental X-Ray Units
G. Inspection of Protective Garments
H. Response to the Reported Problems
I. Visual Inspection
VII.
Diagnostic Equipment
A. Registration of diagnostic units
B. Acquisition of new diagnostic units
C. Disposal of diagnostic units
IX. Important Phone Numbers
I. Proper Operating Procedures
A. Proper Operating Procedures for
Radiographic Units
1. Limit the x-ray primary beam to the smallest area possible
consistent with the objectives of the clinical examination.
2. Align the x-ray beam properly with the patient and the image receptor.
3. Remain behind a protective barrier (i.e., a leaded glass wall or a leaded
door) during the entire radiographic exposure and observe the patient during
the exposure from this protected area.
4. NCRP Report 102, 3.4 (h): "No person should routinely hold patients
during diagnostic examinations. When a patient must be held in position
for radiography, a mechanical supporting or restraining device should be
used. Pregnant women or persons under the age of 18 years should not be
permitted to hold patients. If a patient must be held by someone, that individual
shall be protected with appropriate shielding devices such as protective
aprons and gloves. Positioning should be arranged so that no part of the
holder's torso, even if covered by protective clothing, will be struck by
the useful beam and so that the holder's body is as far as possible from
the useful beam."
5. Do not use expired radiographic films; protect unprocessed film adequately.
6. Provide protective garments (lead aprons and shielding) for all individuals
whose presence is necessary during the radiographic exposure.
B. Proper Operating Procedures for Fluoroscopic
Units
During fluoroscopic examinations of children at CHOP, the parents
often desire to be present in the room. The radiologist or authorized personnel
should discourage the parents from being present. If the presence of parents
is necessary, they should stand as far from the fluoroscope as possible,
and they should wear protective garments (lead aprons, thyroid shield, etc.).
1. As in a radiographic procedure, use the smallest possible
beam area to reduce patient exposure and scatter radiation.
2. Perform visual observation of the alignment of the image intensifier,
x-ray tube, and the patient prior to the initiation of a fluoroscopy procedure.
3. Minimize fluoroscopic doses by reducing the fluoroscopic time used. Fluoroscopic
time, of course, varies with different patients, the type of the examination,
and the complexity of the clinical study. Perform quarterly outputs for
Entrance Skin Exposure (ESE) rates for all fluoroscopes. Post the ESE rates
in the fluoroscopy room for reference by physician/radiologist.
4. Operators should use the timing device to indicate a preset time, which
will serve as a reminder to keep it as short as possible. According to State
regulations, the predetermined time should not exceed 5 minutes (PA 221.38).
5. Use the shortest possible distance from the image intensifier to the
patient. For fluoroscopes that are equipped with AEC; AEC operations provide
for automatic compensation so that when longer distances are used, a higher
radiation dose is given to the patient.
6. The fluoroscopist should wear a thyroid shield, leaded gloves, and glasses
to reduce exposure to the thyroid, extremities, and eyes.
C. Proper Operating Procedures for Mobile Diagnostic
Units
If proper care is not taken, mobile equipment has a greater
potential than standard diagnostic equipment for unnecessary radiation exposure
of personnel and patients. In mobile units, variations can occur in exposures
due to line voltage fluctuations which can affect the output of the machine.
Rechargeable units & capacitor discharge units are not affected.
1. If possible, stand at least 2 meters away from the tube head and the
patient (PA 221.47 c(2)). Distance is often the best possible protection
from radiation.
2. It is important that only individuals necessary for the diagnostic examination
be in the vicinity. Other individuals who are required to remain in the
room should wear protective clothing or should be located behind a protective
shield.
3. Wear protective garments .
4. In mobile or portable radiographic systems, ensure that the source to
skin distance is not less than 30 centimeters. [PA 221.42.c]
5. Mobile fluoroscopes shall provide image intensification (PA 221.40).
D. Proper Operating Procedures for Computed
Tomography Units
1. Use a scan plane position device, such as a light or lights,
to indicate directly or indirectly the position of the slice plane(s) on
the patient within 2 mm (NCRP 3.9.1(e)).
2. Observe the patient during the CT exposures.
3. Only individuals whose presence is necessary should be in the CT x-ray
room during exposures. All such individuals should be protected with leaded
aprons and/or portable shields.
E. Proper Operating Procedures for Dental Units
1. Only persons required for a radiographic procedure should
be in the radiographic room during exposure. All persons must be adequately
protected by protective garments (NCRP 35,4.4).
2. Align the x-ray beam and film very carefully with the area to be radiographed.
3. Stand behind protective barrier and observe the patient during the dental
exposure.
4. Neither the operator nor the assistant shall hold the film in place for
the patient during the exposure. Use the film holder devices to hold the
film in place.
II. Techniques of External Radiation Protection
Control radiation exposure levels via four basic methods:
A. Maximize the distance from radiation source.
B. Minimize the radiation exposure time.
C. Shield the radiation source properly.
D. Shield patients and personnel.
III. Shielding
A. Patient Shielding
"Sensitive body organs (e.g., lens of eye, gonads) should
be shielded whenever they are likely to be exposed to the useful beam provided
that such shielding does not eliminate useful diagnostic information or
proper treatment. Shielding should never be used as a substitute for beam
collimation"(NCRP Report 102, Section 2.2 (d)).
Gonadal Shielding
Shield the gonads with at least 0.25 mm of lead equivalence during diagnostic
procedures in which gonads are in the useful beam, except for cases in which
this would interfere with the diagnostic procedure.
B. Personnel Shielding: personnel who remain
in the room during examinations must be protected by proper shielding.
1. All personnel in the room during an exposure should wear
leaded aprons .
2. Personnel who are likely to be exposed to high levels of scattered radiation
to the thyroid during any procedure should wear thyroid shields.
3. Leaded glasses can greatly reduce the exposure of eye lenses to scattered
radiation in fluoroscopy, especially for physicians.
4. Any person who must have his or her hand near the primary beam (as in
cases in which no other means is available to immobilize a patient) should
wear leaded gloves to reduce exposure of the extremities.
C. Structural Shielding: each radiographic room
has been designed with sufficient shielding in the walls to provide protection
to anyone outside of the room. Do not tamper with the integrity of the shielded
walls. If any personnel notice structural changes, such as holes drilled
into walls, Radiation Safety should be notified as soon as possible.
IV. Pregnant Patient and Pregnant Worker
Policy
A. Patient
Special consideration must be given to the protection of the
embryo or fetus of women known to be, or potentially, pregnant. A patient
of childbearing age should be questioned to ascertain the likelihood of
pregnancy. If the patient is found to be pregnant or likely to be pregnant,
the physician or radiologist should be consulted to decide whether this
radiation dose to the patient is justified.
B. Personnel
An employee should contact the Radiation Safety Office either
directly or through a supervisor when she knows or suspects that she is
pregnant. If for personal reasons an employee does not wish to disclose
pregnancy to her supervisor, confidential disclosures can be made directly
to the Radiation Safety Office.
V. Badging and Dosimetry Policy
A. Personnel Dosimeter Policy
The Radiation Safety Office (RSO) uses personnel monitoring
to identify inadequate or improper radiation safety practices and potentially
serious radiation exposure situations. The RSO will issue proper personnel
dosimeters when evaluation of equipment reveals that the radiation dose
to personnel could potentially be larger than ALARA limits per calendar
quarter to the whole body (125 mrem).
Radiation workers are monitored by the University of Pennsylvania Radiation
Safety Office. All radiation workers must conscientiously wear the radiation
monitoring devices provided by the RSO. These devices may include:
1. A Film Badge to monitor doses to the whole body
2. A Ring Dosimeter to monitor extremity doses
3. A TLD badge for whole body measurements
B. Dosimeter Placement
Interpretation of the measured dose depends on the placement
of the dosimeter. All personnel must wear their dosimeters correctly. The
following list indicates where the dosimeters are to be worn:
1. Film Badge with no TLD Badge: wear the film badge above any protective
clothing at collar level.
2. Film Badge with a TLD Badge: wear the film badge under the lead apron
and the TLD above the apron at collar level.
3. Ring Dosimeter: wear ring dosimeters so that the employee's name faces
outwards.
Do not expose personnel monitoring devices to extreme heat or humidity.
They are screened monthly. If any dosimeter has received a dose higher than
the values shown below, the employee will be notified and the reason for
the high reading will be investigated. Measures will be taken to keep radiation
doses below these limits whenever possible:
RADIOLOGY CARDIAC CATHETERIZATION
WHOLE BODY 125 mrem/calendar quarter
EXTREMITIES 1875 mrem/calendar quarter 300 mrem/calendar quarter
2000 mrem/calendar quarter
All personnel must return all monitoring devices promptly every month.
C. Dose Reports
The Radiation Safety Office sends dose summary reports on a
monthly and an annual basis. Post these reports in a conspicuous location
in the department.
D. Pick-up and Drop-off of Dosimeters
The Radiation Safety Office delivers new dosimeters to each
department on the last Wednesday of each month. Each group should have one
person who is responsible for the distribution of dosimeters. The old dosimeters
should be returned to the designated personnel. The Radiation Safety Office
will pick these up by the first Wednesday of the following month for analysis.
VI. Quality Assurance Program
The quality assurance program will ensure that doses to patients
are in accordance with the standards of good practice.
A. Reject-repeat analysis program
The analysis of the rejected radiographs provides information
about the different aspects of radiological imaging.
B. Film processors
Test each film processor daily for temperature, contrast, density,
and speed. A graph of results of daily measurements will show deviation
from normal behavior.
C. Radiographic x-ray units
Annually test the beam defining system, focal spot size, accuracy
of field size, x-ray field and field alignment, and SID calibration. The
radiation safety characteristics from the unit should comply with NCRP report
NO. 33 as well as FDA and PA State regulations. Evaluate generator performance
by testing the accuracy of the tube potential, timer, and tube current linearity,
and the reproducibility of the x-ray output. In the absence of the mAs indicator,
evaluate the phototimer response. Check the beam quality. Record and the
measurement and compare it to that of the entrance skin exposure for typical
patient examinations.
D. Fluoroscopic x-ray units
Test the exposure rate from the unit under both manual and automatic
operation with simulated patient conditions. The maximum exposure rate should
be less than 10 R/min. Test the high- and low- contrast resolution of the
image intensifier. Measure the half-value layer, effective focal spot size,
and minimum SSD. Test the entrance skin exposure during cine procedures
for cardiology equipment. Measure the Entrance-Skin exposure rates for all
fluoroscopes in each quarter of the calendar year. The radiation safety
characteristics of the machine must meet the recommendations of NCRP Report
NO. 33, as well as FDA and PA state regulations. Make exposure measurements
of scattered radiation under simulated patient conditions.
E. Computed tomographic x-ray units
Test the resolution of the imaging system for high and low contrast.
Measure the patient dose with a phantom under simulated patient conditions.
Using contrast, calculate the noise level. Check the accuracy of tube current
and take a measurement of the scatter radiation. The radiation safety characteristics
of the machine must meet the recommendation of NCRP Report NO. 33, as well
as FDA and PA state regulations.
F. Dental x-ray units
Annually measure focal spot size, collimation, field alignment,
and SSD. Evaluate generator performance by testing the accuracy of the tube
potential, timer, and tube current linearity, as well as the reproducibility
of the x-ray output. Check the beam quality. Record the entrance skin exposure
for a typical patient for typical patient examination and compare it with
the measured doses. The radiation safety characteristics from the unit should
comply with NCRP recommendation report NO. 33, as well as FDA and PA State
regulations. In addition, perform a follow-up survey on all units with major
non-compliance problems. Do a follow-up to ensure that required repairs
are being performed.
G. Inspection of protective garments
Perform an annual inspection of the shielding garments, such
as lead aprons and lead gloves.
H. Response to reported problems
Make all repairs on the units as soon as possible. Send repair
documentation to authorized personnel upon completion of the repairs.
I. Visual inspection
Report all conspicuous problems with energized equipment or
with shielded rooms, as well as any other safety problems observed by personnel
to the Radiation Safety Office immediately.
VII. Diagnostic Equipment
A. Registration of Diagnostic Units
According to Section 216.2 of 25 PA Code, Chapter 216, all radiation-producing
equipment must be registered with the Department of Environmental Resources.
This registration is currently done by the RSO on an annual basis. The RSO
maintains a listing of all units currently registered.
B. Acquisition of New Diagnostic Units
It is the responsibility of clinical personnel to notify the
RSO upon acquisition of any new diagnostic equipment. Authorized RSO personnel
will conduct a radiation safety survey on all new units prior to use.
C. Disposal of Diagnostic Units
Clinical personnel must notify the RSO of any diagnostic equipment
intended for disposal. The RSO will work in cooperation with the Environmental
Health & Safety Office to ensure safe disposal of diagnostic units.
VIII. Important Phone Numbers
Radiation Safety Office of the University of Pennsylvania: 898-7187
CHOP Environmental Health & Safety Office: 590-3872